Synthesis and Characterization of a Biodegradable Polymer

Introduction

Poly(alkylene oxide) polymers are biodegradable and have many important applications, from chemical and biological synthesis to electrolytes for solid state batteries. The polymer may be synthesized from poly(ethylene glycol) (PEG), which is inexpensive and commercially available in average molecular weights from 200 to 10,000. When PEG undergoes a Williamson Condensation reaction under nitrogen and in the presence of KOH in dichloromethane, an unbranched poly(ethylene oxide)is formed (Scheme 1).1,2 For example, PEG400 results in a polymer with average molecular weight of 50,000. After synthesis, this polymer may be characterized by NMR, thin film IR, mechanical load testing, impedance, DSC and SEC. Its mechanical properties may also be compared to other common polymers.


Scheme 1. Polymerization of polyethylene glycol, for example PEG400, where n = 8 to 9.


Synthesis and Characterization

Lab #1

CAUTION: KOH IS CORROSIVE, WEAR GLOVES AT ALL TIMES.

At beginning of the lab, place the glassware to be used in this lab period in the gray oven. The reaction is carried out in the nitrogen-purged glovebox. Finely grind 14.0 g of KOH to a powder in a mortar and pestle. Only when you are ready, remove a 100 ml flask from the oven (use gloves or paper towels), and weigh by difference 12.5 g of KOH into the flask. Add a stir bar to the flask and immediately place into the glove box, along with a teflon spatula and two 25 ml graduated cylinders from the gray oven. Immediately close and purge the antechamber at least twice, as per the posted instructions. From inside the box, retrieve the items, close the antechamber, then purge the box at least twice.

Place a small cold bath (dry ice in acetone, or ice, introduced into the box via a sealed jar) on the magnetic stir plate. Add 12.5 ml of anhydrous dichloromethane (reagent already in the glove box) to the beaker with KOH. Mix the contents manually with the spatula until homogenous, then clamp in place in the ice bath above the stir plate and begin the stirrer. Slowly add 12.5 ml of the viscous poly(ethylene glycol) 400 (reagent also already in the glove box). The reaction is initially exothermic. When the addition of PEG400 is complete, the ice bath is removed from the glove box. The reaction mixture is stirred until it becomes viscous and the stirrer has difficulty stirring (approximately 30 to 60 minutes). An additional 12.5 ml CH2Cl2 is then added. The flask is covered with parafilm, labeled and kept in the glove box with positive nitrogen pressure until the next lab period.

Lab #2

The excess CH2Cl2 is removed using a vacuum line equipped with a nitrogen-cooled trap or a rotovap. Alternatively, heat gently (less than 60°C) in a sand bath in the glove box until dry. The flask is removed from the glove box and the polymer removed from the flask (cleanup of the flask is easy since the polymer is water soluble). Cut the polymer into small pieces and dissolve in ~ 70 ml distilled water in a clean beaker. Mix the slurry with a stir plate. The solution is poured into four dialysis tubes (MW cutoff = 5000 Daltons). The dialysis tubing is cut to approximately the height of the pipette washer, and one end is tied off. Fill the dialysis tubing approximately two-thirds with the aqueous polymer solution. A small funnel may be used for filling. The tubing is kept wet while filling. The ends of the tubing are folded over and secured with the orange clips that can be labeled. When the tubes are clipped, they are placed in a pipet washer with constant distilled water exchange until the next lab period.

Lab #3

Examine the high molecular weight polymer in the dialysis tubing. If there are particles in the aqueous solution, gravity filter through glass wool into a 125 ml Erlenmeyer flask equipped with a magnetic stir bar and gas inlet, weighing the empty flask, bar and inlet apparatus first. Distill away excess water at reduced pressure (again using a vacuum line or rotovap) and dry by heating under vacuum (vacuum line or use a hot water bath--do not boil--and pull vacuum with an aspirator) for at least two hours. It may take longer to remove all of the water. Place the apparatus in the glove box and weigh by difference the mass of product. The glass transition temperature of this polymer is -60°C and its melting temperature is 10°C.

Lab #4 and 5

Place a small amount of the polymer in a vial and remove from the glove box. Dissolve a small amount of polymer in deuterated chloroform and run the 1H NMR spectrum in S2-G14. You need to coordinate with the NMR specialist, Dr. Schulte one lab period in advance so that he can run your NMR sample. Also obtain the IR spectrum by dissolving a small amount of the polymer in acetonitrile in the fumehood and make a thin film on shrink wrap by dropping the polymer solution onto the wrap and allowing to air-dry. Alternatively, use NaCl plates with either the pure polymer or Nujol mull.

Finally, evaluate the strength of the polymer on the mechanical load testing equipment. Make a solid piece of the polymer be heating on a hot plate into a mold, or heat the polymer into one piece and cut into shape with an industrial blade. Your instructor will mount the sample on the Instron equipment and collect the data. Collect the data for several other polymers that are also supplied.


Lab Report

Based upon the integration in your 1H NMR, calculate the ratio of methoxy to ethoxy peaks in your co-block polymer. How does this NMR compare with the NMR of PEG400? Note: there are no methoxy peaks in PEG400, while there are terminal hydroxymethylene (CH2OH) peaks in both PEG400 and the co-block polymer. Do not compare integrations between spectra, since they are relative to the specific sample only.

Assign the significant peaks in your IR to specific organic functionalities, paying particular attention to the peaks from 1500 to 600 cm-1. How does the experimental IR compare with the reference IR?2

Compare the polymer strength with known polymers, both like and unlike the polymer you synthesized.

Acknowledgments

The mechanical testing portion of this experiment was supported by a grant from the NSF-DUE-CCLI-A&I (0310454, PI: Professor Stevens).

References

1. Lemmon, J. P.; Kohnert, R. L.; Lerner, M. M. Macromolecules 1993, 26, 2767- 2770.

2. Lemmon, J. P.; Lerner, M. M. Macromolecules 1992, 25, 2907-2909.